Background
Covered stent correction of sinus venosus ASDs (SVASD) is a relatively new technique. Challenges include anchoring a sufficiently long stent in a nonstenotic superior vena cava (SVC) and expanding the stent at the wider SVC‐RA junction without obstructing the anomalous right upper pulmonary vein (RUPV). The 10‐zig covered Cheatham‐platinum (CCP) stent has the advantage of being available in lengths of 5–11 cm and dilatable to 34 mm in diameter.
Methods
An international registry reviewed the outcomes of 10‐zig CCP stents in 75 patients aged 11.4–75.9 years (median 45.4) from March 2016. Additional stents were used to anchor the stent in the SVC or close residual shunts in 33/75. An additional stent was placed in 4/5 (80%) with 5/5.5 cm CCPs, 18/29 (62%) with 6 cm CCPs, 5/18 (28%) with 7 cm CCPs, 5/22 (23%) with 7.5/8 cm CCPs and 0/1 with an 11 cm CCP. A “protective …