作者
Lorenz Birnbacher, Marian Willner, Astrid Velroyen, Mathias Marschner, Alexander Hipp, Jan Meiser, Frieder Koch, Tobias Schröter, Danays Kunka, Jürgen Mohr, Franz Pfeiffer, Julia Herzen
发表日期
2016/4/4
期刊
Scientific reports
卷号
6
期号
1
页码范围
24022
出版商
Nature Publishing Group UK
简介
The possibility to perform high-sensitivity X-ray phase-contrast imaging with laboratory grating-based phase-contrast computed tomography (gbPC-CT) setups is of great interest for a broad range of high-resolution biomedical applications. However, achieving high sensitivity with laboratory gbPC-CT setups still poses a challenge because several factors such as the reduced flux, the polychromaticity of the spectrum and the limited coherence of the X-ray source reduce the performance of laboratory gbPC-CT in comparison to gbPC-CT at synchrotron facilities. In this work, we present our laboratory X-ray Talbot-Lau interferometry setup operating at 40 kVp and describe how we achieve the high sensitivity yet unrivalled by any other laboratory X-ray phase-contrast technique. We provide the angular sensitivity expressed via the minimum resolvable refraction angle both in theory and experiment and compare our data …
引用总数