作者
Peiying Liu, Feng Xu, Hanzhang Lu
发表日期
2012/4/19
期刊
Magnetic Resonance in Medicine
卷号
69
期号
3
页码范围
675-681
出版商
Wiley Subscription Services, Inc., A Wiley Company
简介
Cerebral metabolic rate of oxygen (CMRO2) is an important index of tissue viability and brain function, but this parameter cannot yet be measured routinely on clinical scanners. Recently, a noninvasive technique was proposed which estimates global CMRO2 by concomitantly measuring oxygen‐extraction‐fraction using T2‐relaxation‐under‐spin‐tagging MRI and pulse oximetry, and cerebral‐blood‐flow using phase‐contrast MRI. This study sought to establish a standard acquisition procedure for this technique and to evaluate its test–retest reproducibility in healthy subjects. Each subject was examined in five sessions and each session included two measurements. Intrasession, intersession, and intersubject coefficients of variation for CMRO2 were found to be 3.84 ± 1.44% (N = 7, mean ± standard deviation), 6.59 ± 1.56%, and 8.80% respectively. These reproducibility values were comparable or slightly …
引用总数
2012201320142015201620172018201920202021202220232024131291112131010101163