作者
Dale L Bailey, Brian F Hutton, Paul J Walker
发表日期
1987/5/1
期刊
Journal of Nuclear Medicine
卷号
28
期号
5
页码范围
844-851
出版商
Society of Nuclear Medicine
简介
A methodis proposedto simultaneouslyrecordsinglephotonemissionand transmission tomographic (SPECT) studiesto producea mapof attenuationcoefficients (@@) for the body. A dual radionuclide SPECT acquisition is performed with a transmission source attached to a rotating gamma camera of lower energy than the emission radionuclide. Scatter from the emissionsourceinto the transmissionwindow is removedby subtractingthe predicted scatterdistribution. Thetransmissionimageisthenreconstructedtoyieldthemapof attenuationcoefficientsfor anatomicdisplayor attenuationcorrectionpurposes. Experimental work has shown that the method can accurately derive@ values to ±2.5% in both phantom andpatientstudies, withoutincreasingacquisitiontime. Preliminaryattenuationcorrection experiments have demonstrated an accuracy of better than 5% for estimated activity.
引用总数
学术搜索中的文章
DL Bailey, BF Hutton, PJ Walker - Journal of Nuclear Medicine, 1987