作者
Yusuf Samet Yaras, Sarp Satir, Cagla Ozsoy, Rajiv Ramasawmy, Adrienne E Campbell-Washburn, Robert J Lederman, Ozgur Kocaturk, F Levent Degertekin
发表日期
2018/9/5
期刊
IEEE Transactions on Biomedical Engineering
卷号
66
期号
4
页码范围
1148-1154
出版商
IEEE
简介
Objective
The objective of this paper is to introduce an acousto-optic optical fiber sensor for tracking catheter position during interventional magnetic resonance imaging (MRI) to overcome RF induced heating of active markers.
Methods
The sensor uses a miniature coil coupled to a piezoelectric transducer, which is in turn mechanically connected to an optical fiber. The piezoelectric transducer converts the RF signal to acoustic waves in the optical fiber over a region including a fiber Bragg grating (FBG). The elastic waves in the fiber modulates the FBG geometry and hence the reflected light in the optical fiber. Since the coil is much smaller than the RF wavelength and the signal is transmitted on the dielectric optical fiber, the sensor effectively reduces RF induced heating risk. Proof of concept prototypes of the sensor are implemented using commercially available piezoelectric transducers and optical fibers with …
引用总数
201920202021202220232024237252
学术搜索中的文章
YS Yaras, S Satir, C Ozsoy, R Ramasawmy… - IEEE Transactions on Biomedical Engineering, 2018