Mohsen Agharazii, Isabelle Plamondon, Marcel Lebel, Pierre Douville, Simon Desmeules
发表日期
2005/6/1
期刊
Nephrology Dialysis Transplantation
卷号
20
期号
6
页码范围
1238-1240
出版商
Oxford University Press
简介
Background. Although most catheter problems in haemodialysis are related to infection or clotting, bleeding associated with the heparin lock is of clinical importance especially during peri-operative conditions. The objective of this in vitro study is to estimate the volume of heparin that may leak into the circulation immediately after performing a catheter lock.
Methods. Different volumes (ml) of a dextrose solution were used to perform a catheter lock on haemodialysis catheters. The tip of the catheter was placed in a test tube containing water for a pre-specified period. The final concentrations of dextrose in the test tube were used to determine the volume of solution that leaked from the catheter.
Results. When the total lumen volume was filled, the catheter leak was estimated to be 0.59±0.03 and 0.71±0.04 ml after 15 and 25 s, respectively. There was a continuous leak of 1.23±0.41, 2 …