作者
F Roellinghoff, A Benilov, D Dauvergne, G Dedes, N Freud, G Janssens, J Krimmer, Jean Michel Létang, M Pinto, D Prieels, C Ray, J Smeets, F Stichelbaut, E Testa
发表日期
2014/2/20
期刊
Physics in Medicine & Biology
卷号
59
期号
5
页码范围
1327
出版商
IOP Publishing
简介
Prompt-gamma profile was measured at WPE-Essen using 160 MeV protons impinging a movable PMMA target. A single collimated detector was used with time-of-flight (TOF) to reduce the background due to neutrons. The target entrance rise and the Bragg peak falloff retrieval precision was determined as a function of incident proton number by a fitting procedure using independent data sets. Assuming improved sensitivity of this camera design by using a greater number of detectors, retrieval precisions of 1 to 2 mm (rms) are expected for a clinical pencil beam. TOF improves the contrast-to-noise ratio and the performance of the method significantly.
引用总数
2014201520162017201820192020202120222023202461082125107582
学术搜索中的文章