作者
Ali Kiapour, Tunc Oktenoglu, Ali F Ozer, Vijay K Goel
简介
Currently, rigid fixation systems are the gold standard for degenerative disk disease treatment. Dynamic fixation systems have been proposed as alternatives for the treatment of a variety of spinal disorders. These systems address the main drawbacks of traditional rigid fixation systems, such as adjacent segment degeneration and instrumentation failure. Pedicle-screw-based dynamic stabilization (PDS) is one type of these alternative systems. The aim of this study was to simulate the biomechanical effect of a novel posterior dynamic stabilization system, which is comprised of dynamic (hinged) screws interconnected with a coiled, spring-based dynamic rod (DSDR), and compare it to semirigid (DSRR and RSRR) and rigid stabilization (RSRR) systems. A validated finite element (FE) model of L1-S1 was used to quantify the biomechanical parameters of the spine, such as range of motion, intradiskal pressure, stresses and facet loads after single-level instrumentation with different posterior stabilization systems. The results obtained from in vitro experimental intact and instrumented spines were used to validate the FE model, and the validated model was then used to compare the biomechanical effects of different fixation and stabilization constructs with intact under a hybrid loading protocol. The segmental motion at L4–L5 increased by 9.5% and 16.3% in flexion and left rotation, respectively, in DSDR with respect to the intact spine, whereas it was reduced by 6.4% and 10.9% in extension and left-bending loads, respectively. After instrumentation-induced intradiskal pressure at adjacent segments, L3-L4 and L5-S1 became less than the intact in …