作者
John T Capo, Eitan Melamed, Benhoor Shamian, Scott R Hadley, Whitney Ng Lai, Kenny Gerszberg, Steven Rivero, Paolo Caravaggi
发表日期
2014/10/1
期刊
The Journal of hand surgery
卷号
39
期号
10
页码范围
1971-1977
出版商
WB Saunders
简介
Purpose
To determine in a cadaver model which, among 5 fixation methods for proximal interphalangeal (PIP) joint arthrodesis, has the greatest stiffness.
Methods
Thirty-five cadaver digits were randomly assigned to 1 of 5 fixation groups: oblique K-wire with coronal intraosseous wiring, tension-band wire (TBW), dorsal plate, intramedullary linked screw (IMS), and 90/90 wiring (90/90W). Testing was done by applying bending moments to the PIP joint in the sagittal and frontal planes. The force/displacement curves were used to estimate the stiffness of each construct. Ultimate strength was determined by loading to failure in extension.
Results
The IMS had significantly higher stiffness than all wiring constructs in all planes of motion and significantly greater stiffness in extension than the dorsal plate. The IMS stiffness exceeded 10 N/mm across all bending directions and showed an ultimate strength of 21 N. The plate …
引用总数
2015201620172018201920202021202220232024245462461
学术搜索中的文章