作者
Annegret H Dahlmann-Noor, Kalliopi Vrotsou, Vasileios Kostakis, Janet Brown, Jayne Heath, Abigail Iron, Stuart McGill, Anthony J Vivian
发表日期
2009/3/1
期刊
British journal of ophthalmology
卷号
93
期号
3
页码范围
342-345
出版商
BMJ Publishing Group Ltd
简介
Background/aims
To evaluate a new autorefractor, the Plusoptix Vision Screener (PVS), as a screening tool to detect risk factors for amblyopia by comparing it with gold-standard orthoptic vision screening in children.
Methods
Community-based screening study including 288 children age 4–7 years who were screened with the PVS and by orthoptic assessment (distance acuity, cover test, extraocular movements, 20 PD prism test, Lang stereotest). Follow-up comprehensive eye examination of screening-positive children included manual cycloplegic retinoscopy.
Results
Testability was high for both methods. Orthoptic screening identified 36 children with reduced vision and/or factors associated with amblyopia (referral rate 12.5%). The PVS identified 16 children with potential vision problems (referral rate 5.6%), indicating only moderate sensitivity (44%; 95% CI 27.9 to 61.9%), but high specificity (100%; 95% CI 98.5 to …
引用总数
20092010201120122013201420152016201720182019202020212022202320242353465885375223
学术搜索中的文章