Angeliki Ntineri, Sofia Prapa, Ioanna Bountzona, Ariadni Menti, George S Stergiou
发表日期
2021/4/1
期刊
Blood Pressure Monitoring
卷号
26
期号
2
页码范围
146-148
出版商
LWW
简介
Objective
The aim of this study was to evaluate the accuracy of the single upper-arm cuff oscillometric blood pressure (BP) monitor InBody BPBIO750 developed for self-measurement by adults in public spaces (kiosk) according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2: 2018).
Methods
Subjects were recruited to fulfil the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method.
Results
A total of 102 subjects were recruited and 85 were analyzed [mean age 56.7±15.4 (SD) years, 40 men, arm circumference 32.3±5.3 cm, range 22–42 cm]. For validation criterion 1, the mean±SD of the differences between the test device and reference BP readings was 2.2±6 …