作者
Teryn R Roberts, Antoine Persello, George T Harea, Else M Vedula, Brett C Isenberg, Yanyi Zang, Jose Santos, Jeffrey T Borenstein, Andriy I Batchinsky
发表日期
2024/6/1
期刊
ASAIO Journal
卷号
70
期号
6
页码范围
535-544
出版商
LWW
简介
Microfluidic membrane oxygenators are designed to mimic branching vasculature of the native lung during extracorporeal lung support. To date, scaling of such devices to achieve clinically relevant blood flow and lung support has been a limitation. We evaluated a novel multilayer microfluidic blood oxygenator (BLOx) capable of supporting 750–800 ml/min blood flow versus a standard hollow fiber membrane oxygenator (HFMO) in vivo during veno-venous extracorporeal life support for 24 hours in anesthetized, mechanically ventilated uninjured swine (n= 3/group). The objective was to assess feasibility, safety, and biocompatibility. Circuits remained patent and operated with stable pressures throughout 24 hours. No group differences in vital signs or evidence of end-organ damage occurred. No change in plasma free hemoglobin and von Willebrand factor multimer size distribution were observed. Platelet count …