作者
Todd J Brinton, Bruno Cotter, Mala T Kailasam, David L Brown, Shiu-Shin Chio, Daniel T O’Connor, Anthony N DeMaria
发表日期
1997/8/1
期刊
The American journal of cardiology
卷号
80
期号
3
页码范围
323-330
出版商
Excerpta Medica
简介
The ability not only to record automated systolic and diastolic pressure, but also to derive measurements of the rate of pressure change during the cardiac cycle, would have great potential clinical value. A new method has been developed to obtain pressure measurements at 20-ms intervals by oscillometric cuff signal pattern recognition. Derivation of noninvasive pressure measurements is based on a T tube aorta and straight tube brachial artery, and assumes that the systolic phase of the suprasystolic cuff signal and the diastolic phase of the subdiastolic cuff signal most closely approximate systolic and diastolic aortic pressures, respectively. Arterial pressures obtained by this method were compared with simultaneous invasive measurements from the thoracic aorta in 36 patients. Good agreement was observed between noninvasive and invasive methods for systolic (146 ± 4 vs 145 ± 5 mm Hg), diastolic (80 ± 2 vs …