作者
Pejman Rowshanfarzad, Mahsheed Sabet, Daryl J O'Connor, Peter M McCowan, Boyd MC McCurdy, Peter B Greer
发表日期
2012/2
期刊
Medical physics
卷号
39
期号
2
页码范围
623-635
出版商
American Association of Physicists in Medicine
简介
Purpose
: Electronic portal imaging devices (EPIDs) have been studied and used for pretreatment and in‐vivo dosimetry applications for many years. The application of EPIDs for dosimetry in arc treatments requires accurate characterization of the mechanical sag of the EPID and gantry during rotation. Several studies have investigated the effects of gravity on the sag of these systems but each have limitations. In this study, an easy experiment setup and accurate algorithm have been introduced to characterize and correct for the effect of EPID and gantry sag during arc delivery.
Methods
: Three metallic ball bearings were used as markers in the beam: two of them fixed to the gantry head and the third positioned at the isocenter. EPID images were acquired during a 360° gantry rotation in cine imaging mode. The markers were tracked in EPID images and a robust in‐house developedMATLAB code was used to …