作者
Laurence J Russell, Tamara Dodd, Daniel Kendall, Amber Lazenbury, Abigail Leggett, Sophie Payton-Haines, Liudi Jiang, Davide Filingeri, Peter R Worsley
发表日期
2024/2/1
期刊
Clinical Biomechanics
卷号
112
页码范围
106178
出版商
Elsevier
简介
Background
Cervical collars restrict cervical spine movement to minimise the risk of spinal cord injury. Collars apply mechanical loading to the skin putting it at risk of skin damage. Indeed, cervical collar-related pressure ulcers are unacceptably prevalent, especially at the occiput, mandibles, and chin. Collar design and fit are often key considerations for prevention.
Methods
This comprehensive study evaluated four commercial prehospital and acute care cervical collars. Pressure, microclimate, transepidermal water loss and skin hydration were measured at the interface between the device and the skin. Range of motion restriction was measured to evaluate effective immobilisation. Head, neck, and shoulder morphology was evaluated using three-dimensional scans.
Findings
The occiput experienced significantly higher interface pressures than the chin and mandibles for most collar designs. Interface pressure at the …