作者
Sierra Farris, Monique Lillian Giroux
发表日期
2010/1
期刊
Neuromodulation: Technology at the Neural Interface
卷号
13
期号
1
页码范围
65-67
出版商
Blackwell Publishing Asia
简介
Objectives. We present a case report of a patient with complete lead wire fracture with approximately 3‐mm separation of the wire fragments that had three electrode impedance measurements and therapy impedance measurement that would be consistent with an intact wire. Materials and Methods. Retrospective chart review. Results. The patient experienced abrupt worsening of gait freezing while tremor control remained stable. Despite normal therapeutic impedance measurement, clinical suspicion for a hardware problem remained due to her description of brief dystonic hand posturing witnessed during impedance measurements in the clinic. This led to further diagnostic evaluation and eventual surgical replacement of the lead wire. Conclusions. Impedance measurements are not absolutely reliable for the detection of hardware malfunction in a patient with a Kinetra neurostimulator.