G Cao, YZ Lee, R Peng, Z Liu, R Rajaram, X Calderon-Colon, L An, P Wang, T Phan, S Sultana, DS Lalush, JP Lu, O Zhou
发表日期
2009/3/25
期刊
Physics in Medicine & Biology
卷号
54
期号
8
页码范围
2323
出版商
IOP Publishing
简介
Current commercial micro-CT scanners have the capability of imaging objects ex vivo with high spatial resolution, but performing in vivo micro-CT on free-breathing small animals is still challenging because their physiological motions are non-periodic and much faster than those of humans. In this paper, we present a prototype physiologically gated micro-computed tomography (micro-CT) scanner based on a carbon nanotube field emission micro-focus x-ray source. The novel x-ray source allows x-ray pulses and imaging sequences to be readily synchronized and gated to non-periodic physiological signals from small animals. The system performance is evaluated using phantoms and sacrificed and anesthetized mice. Prospective respiratory-gated micro-CT images of anesthetized free-breathing mice were collected using this scanner at 50 ms temporal resolution and 6.2 lp mm− 1 at 10% system MTF. The high …