作者
Chad A Zarse, James A McAteer, Mark Tann, Andre J Sommer, Samuel C Kim, Ryan F Paterson, Erin K Hatt, James E Lingeman, Andrew P Evan, James C Williams Jr
发表日期
2004/5/1
期刊
Urology
卷号
63
期号
5
页码范围
828-833
出版商
Elsevier
简介
OBJECTIVES
To assess the ability of helical computed tomography (CT) to differentiate regions of known mineral composition in typical, heterogeneous urinary stones. Interest is substantial in the urologic community in using radiologic imaging to determine accurately the composition of urinary calculi. Recent advances in CT make this a viable prospect, but the heterogeneity of most stones is a complicating factor.
METHODS
The ability of micro-CT (a high-resolution laboratory instrument) to identify the mineral composition of stones was confirmed by calibrating micro-CT attenuation values to pure mineral regions of sliced stones using infrared microspectroscopy. Intact human urinary stones were then analyzed by micro-CT, and regions-of-interest of pure mineral were correlated with identical regions-of-interest from quad slice multi-detector row helical CT images. With helical CT, narrow slice widths were used to …