作者
Andrew N Primak, Joel G Fletcher, Terri J Vrtiska, Oleksandr P Dzyubak, John C Lieske, Molly E Jackson, James C Williams Jr, Cynthia H McCollough
发表日期
2007/12/1
期刊
Academic radiology
卷号
14
期号
12
页码范围
1441-1447
出版商
Elsevier
简介
RATIONALE AND OBJECTIVES
To determine the accuracy and sensitivity for dual-energy computed tomography (DECT) discrimination of uric acid (UA) stones from other (non-UA) renal stones in a commercially implemented product.
MATERIALS AND METHODS
Forty human renal stones comprising uric acid (n = 16), hydroxyapatite (n = 8), calcium oxalate (n = 8), and cystine (n = 8) were inserted in four porcine kidneys (10 each) and placed inside a 32-cm water tank anterior to a cadaver spine. Spiral dual-energy scans were obtained on a dual-source, 64-slice computed tomography (CT) system using a clinical protocol and automatic exposure control. Scanning was performed at two different collimations (0.6 mm and 1.2 mm) and within three phantom sizes (medium, large, and extra large) resulting in a total of six image datasets. These datasets were analyzed using the dual-energy software tool available on …
引用总数