Andrew P Santoso, Kwang H Song, Yujiao Qin, Stephen J Gardner, Chang Liu, Indrin J Chetty, Benjamin Movsas, Munther Ajlouni, Ning Wen
发表日期
2016/12
期刊
Radiation Oncology
卷号
11
页码范围
1-11
出版商
BioMed Central
简介
Background
This study investigates the effect of gantry speed on 4DCBCT image quality and dose for the Varian On-Board Imager®.
Methods
A thoracic 4DCBCT protocol was designed using a 125 kVp spectrum. Image quality parameters were evaluated for 4DCBCT acquisition using Catphan® phantom with real-time position management™ system for gantry speeds varying between 1.0 to 6.0°/s. Superior-inferior motion of the phantom was executed using a sinusoidal waveform with five second period. Scans were retrospectively sorted into 4 phases (CBCT-4 ph) and 10 phases (CBCT-10 ph); average 4DCBCT (CBCT-ave), using all image data from the 4DCBCT acquisitions was also evaluated. The 4DCBCT images were evaluated using the following image quality metrics: spatial resolution, contrast-to-noise ratio (CNR), and uniformity index (UI …