Elena Salvagnini, Hilde Bosmans, Lara Struelens, Nicholas W Marshall
发表日期
2012/6
期刊
Medical Physics
卷号
39
期号
6Part1
页码范围
3167-3180
出版商
American Association of Physicists in Medicine
简介
Purpose:
Four different practical methodologies of quantifying scattered radiation for two different digital mammographic systems are compared. The study considered both grid in and grid out geometries for two different antiscatter grid types, a typical linear grid and a cellular grid design. The aim was to find quick and reproducible methods that could be used in place of the beam stop technique.
Methods:
The scatter to primary ratio (SPR) and the scatter fraction (SF) were used to quantify scattered radiation as a function of poly(methyl methacrylate) (PMMA) thickness, grid position, and beam quality. The four scatter estimation methods applied were (1) the beam stop method, (2) a hybrid method that combined measured detector (scatter‐free) modulation transfer function (MTF) data and a Monte Carlo simulation of the scatter point spread function, (3) from the low frequency drop data taken from the system MTF …