Lesley Cockmartin, NW Marshall, Guozhi Zhang, Kim Lemmens, Eman Shaheen, Chantal Van Ongeval, Erik Fredenberg, DR Dance, Elena Salvagnini, Koen Michielsen, Hilde Bosmans
发表日期
2017/1/10
期刊
Physics in Medicine & Biology
卷号
62
期号
3
页码范围
758
出版商
IOP Publishing
简介
This paper introduces and applies a structured phantom with inserted target objects for the comparison of detection performance of digital breast tomosynthesis (DBT) against 2D full field digital mammography (FFDM). The phantom consists of a 48 mm thick breast-shaped polymethyl methacrylate (PMMA) container filled with water and PMMA spheres of different diameters. Three-dimensionally (3D) printed spiculated masses (diameter range: 3.8–9.7 mm) and non-spiculated masses (1.6–6.2 mm) along with microcalcifications (90–250 µm) were inserted as targets.
引用总数
2017201820192020202120222023202431011159993