作者
R Lee MacDonald, Young Lee, Jannie Schasfoort, Hany Soliman, Arjun Sahgal, Mark Ruschin
发表日期
2020/2/1
期刊
International Journal of Radiation Oncology* Biology* Physics
卷号
106
期号
2
页码范围
413-421
出版商
Elsevier
简介
Purpose
The transition from frame-based brain stereotactic radiosurgery (SRS) to frameless delivery is supported by real-time intrafraction monitoring to ensure accurate delivery. The purpose of this study is to characterize these real-time motion traces in a large cohort of patients treated with frameless gated brain SRS and to develop patient-specific predictions of tolerance violations.
Methods and Materials
SRS patients treated on the Gamma Knife Icon were immobilized using a device-specific thermoplastic head mask. A motion marker was fixed to the patient’s nose, with gating and cone beam computed tomography (CBCT)-based corrections to the treatment at excursions from baseline exceeding 1.5 mm. The traces of 1446 fractions were analyzed according to magnitude (932 unique treatment plans for 462 unique individual patients), directional distribution of displacement, and stability. A neural network model …
引用总数
20202021202220232024681161
学术搜索中的文章
RL MacDonald, Y Lee, J Schasfoort, H Soliman… - International Journal of Radiation Oncology* Biology …, 2020