Purpose
This patient study evaluated the use of 3-dimensional (3D) printed bolus for chest wall radiation therapy compared with standard sheet bolus with regard to accuracy of fit, surface dose measured in vivo, and efficiency of patient setup. By alternating bolus type over the course of therapy, each patient served as her own control.
Methods and materials
For 16 patients undergoing chest wall radiation therapy, a custom 5.0 mm thick bolus was designed based on the treatment planning computed tomography scan and 3D printed using polylactic acid. Cone beam computed tomography scanning was used to image and quantify the accuracy of fit of the 2 bolus types with regard to air gaps between the bolus and skin. As a quality assurance measure for the 3D printed bolus, optically stimulated luminescent dosimetry provided in vivo comparison of surface dose at 7 points on the chest wall. Durations of patient setup …