作者
Yoon Young Cho, Sejong Chun, Soo-Youn Lee, Jae Hoon Chung, Hyung-Doo Park, Sun Wook Kim
发表日期
2016/9
期刊
Annals of laboratory medicine
卷号
36
期号
5
页码范围
413
出版商
Korean Society for Laboratory Medicine
简介
Background: Measurement of postoperative serum thyroglobulin (Tg) is important for detecting persistent or recurrent differentiated thyroid cancer. We evaluated the analytic performance of the DxI 800 assay (Beckman Coulter, USA) for serum Tg and anti-thyroglobulin antibodies (TgAbs) in comparison with that of the GAMMA-10 assay (Shinjin Medics Inc., Korea) for serum Tg and RIA-MAT 280 assay (Stratec, Germany) for TgAb. Methods: We prospectively collected blood samples from 99 patients thyroidectomized for thyroid cancer. The functional sensitivity was investigated in standards and human serum. Precision and linearity were evaluated according to the guidelines of the Clinical and Laboratory Standards Institute. The correlation between the two assays was assessed in samples with different Tg ranges.
Results: The functional sensitivity of the DxI 800 assay for serum Tg was between 0.0313 and 0.0625 ng/mL. The total CV was 3.9-5.6% for serum Tg and 5.3-6.9% for serum TgAb. The coefficient of determination (R2) was 1.0 and 0.99 for serum Tg and TgAb, respectively. The cut-offs for serum TgAb were 4.0 IU/mL (DxI 800) and 60.0 IU/mL (RIAMAT 280), and the overall agreement was 68.7%. The correlation between the two assays was excellent; the correlation coefficient was 0.99 and 0.88 for serum Tg and TgAb, respectively.
Conclusions: The DxI 800 is a sensitive assay for serum Tg and TgAb, and the results correlated well with those from the immunoradiometric assays (IRMA). This assay has several advantages over the IRMA and could be considered an alternative test for Tg measurement.
引用总数
201720182019202020212022202320242434111