Mariana Yalon, Akitoshi Inoue, Jamison E Thorne, Yong S Lee, Matthew P Johnson, Andrea Esquivel, Shuai Leng, Cynthia H McCollough, Joel G Fletcher, Prabhakar Shantha Rajiah
发表日期
2024/3
期刊
American Journal of Roentgenology
卷号
222
期号
2
页码范围
e2329778
出版商
American Roentgen Ray Society
简介
BACKGROUND. The higher spatial resolution and image contrast for iodine-containing tissues of photon-counting detector (PCD) CT may address challenges in evaluating small calcified vessels when performing lower extremity CTA by energy-integrating detector (EID) CTA.
OBJECTIVE. The purpose of the study was to compare the evaluation of infrapopliteal vasculature between lower extremity CTA performed using EID CT and PCD CT.
METHODS. This prospective study included 32 patients (mean age, 69.7 ± 11.3 [SD] years; 27 men, five women) who underwent clinically indicated lower extremity EID CTA between April 2021 and March 2022; participants underwent investigational lower extremity PCD CTA later the same day as EID CTA using a reduced IV contrast media dose. Two radiologists independently reviewed examinations in two sessions, each containing a random combination of EID CTA and …