作者
J Rafe Sales, Thomas J Ellis, Joel Gillard, Qi Liu, Joyce C Chen, Bruce Ham, John C Mayberry
发表日期
2008/5/1
期刊
Journal of Trauma and Acute Care Surgery
卷号
64
期号
5
页码范围
1270-1274
出版商
LWW
简介
Background:
A novel rib fracture repair plating system was developed to provide durable fixation with a shorter length than standard systems and thus facilitate minimally invasive repair. We hypothesized that U-plate fixation would be at least equivalent in durability to standard anterior fixation.
Study:
Twenty fresh frozen ribs (10 pairs) from two human cadavers were first tested for intact stiffness (force or deformation). A gap of 5 mm was then created in the middle of each rib with a saw. Each rib was reconstructed with either the U-plate (4.6 cm length, Acute Innovations, LLC, Hillsboro, OR) with four screws or a 2.4-mm anterior locking plate (9.5 cm length, Synthes, Paoli, PA) with six screws. The U-plates were placed on one rib and the anterior plates on the contralateral rib of the paired levels. The reconstructed ribs were cycled 50,000 times with a load of±2N at 1 Hz in a simulation of the repetitive loading of deep …
引用总数
20092010201120122013201420152016201720182019202020212022202320244106638131168515341
学术搜索中的文章
JR Sales, TJ Ellis, J Gillard, Q Liu, JC Chen, B Ham… - Journal of Trauma and Acute Care Surgery, 2008