作者
Markus Weiger, Klaas P Pruessmann, Christoph Leussler, Peter Röschmann, Peter Boesiger
发表日期
2001/3
期刊
Magnetic Resonance in Medicine: An Official Journal of the International Society for Magnetic Resonance in Medicine
卷号
45
期号
3
页码范围
495-504
出版商
John Wiley & Sons, Inc.
简介
In sensitivity encoding (SENSE), the effects of inhomogeneous spatial sensitivity of surface coils are utilized for signal localization in addition to common Fourier encoding using magnetic field gradients. Unlike standard Fourier MRI, SENSE images exhibit an inhomogeneous noise distribution, which crucially depends on the geometrical sensitivity relations of the coils used. Thus, for optimum signal‐to‐noise‐ratio (SNR) and noise homogeneity, specialized coil configurations are called for. In this article we study the implications of SENSE imaging for coil layout by means of simulations and imaging experiments in a phantom and in vivo. New, specific design principles are identified. For SENSE imaging, the elements of a coil array should be smaller than for common phased‐array imaging. Furthermore, adjacent coil elements should not overlap. Based on the findings of initial investigations, a configuration of six …
引用总数
2000200120022003200420052006200720082009201020112012201320142015201620172018201920202021202220232024551222262225372613111326136734531222
学术搜索中的文章
M Weiger, KP Pruessmann, C Leussler, P Röschmann… - Magnetic Resonance in Medicine: An Official Journal of …, 2001