作者
Saman Nazarian, Ariel Roguin, Menekhem M Zviman, Albert C Lardo, Timm L Dickfeld, Hugh Calkins, Robert G Weiss, Ronald D Berger, David A Bluemke, Henry R Halperin
发表日期
2006/9/19
期刊
Circulation
卷号
114
期号
12
页码范围
1277-1284
出版商
Lippincott Williams & Wilkins
简介
Background— Magnetic resonance imaging (MRI) is an important diagnostic modality currently unavailable for millions of patients because of the presence of implantable cardiac devices. We sought to evaluate the diagnostic utility and safety of noncardiac and cardiac MRI at 1.5T using a protocol that incorporates device selection and programming and limits the estimated specific absorption rate of MRI sequences.
Methods and Results— Patients with no imaging alternative and with devices shown to be MRI safe by in vitro phantom and in vivo animal testing were enrolled. Of 55 patients who underwent 68 MRI studies, 31 had a pacemaker, and 24 had an implantable defibrillator. Pacing mode was changed to “asynchronous” for pacemaker-dependent patients and to “demand” for others. Magnet response and tachyarrhythmia functions were disabled. Blood pressure, ECG, oximetry, and symptoms were …