作者
Ariel Roguin, Menekhem M Zviman, Glenn R Meininger, E Rene Rodrigues, Timm M Dickfeld, David A Bluemke, Albert Lardo, Ronald D Berger, Hugh Calkins, Henry R Halperin
发表日期
2004/8/3
期刊
Circulation
卷号
110
期号
5
页码范围
475-482
出版商
Lippincott Williams & Wilkins
简介
Background— MRI has unparalleled soft-tissue imaging capabilities. The presence of devices such as pacemakers and implantable cardioverter/defibrillators (ICDs), however, is historically considered a contraindication to MRI. These devices are now smaller, with less magnetic material and improved electromagnetic interference protection. Our aim was to determine whether these modern systems can be used in an MR environment.
Methods and Results— We tested in vitro and in vivo lead heating, device function, force acting on the device, and image distortion at 1.5 T. Clinical MR protocols and in vivo measurements yielded temperature changes <0.5°C. Older (manufactured before 2000) ICDs were damaged by the MR scans. Newer ICD systems and most pacemakers, however, were not. The maximal force acting on newer devices was <100 g. Modern (manufactured after 2000) ICD systems were …