作者
Sunil Shah, Mohammad Laiquzzaman, Ian Yeung, Xueliang Pan, Cynthia Roberts
发表日期
2009/6/1
期刊
Contact Lens and Anterior Eye
卷号
32
期号
3
页码范围
123-128
出版商
Elsevier
简介
PURPOSE
To compare corneal biomechanical parameters and two measures of intraocular pressure (IOP) in eyes before and after excimer laser refractive surgery, with the Ocular Response Analyser (ORA).
MATERIALS AND METHODS
Eighty normal eyes of 41 patients undergoing excimer laser refractive surgery in Birmingham, U.K. were recruited into three groups: Laser Assisted-Epithelial Keratomileusis (LASEK) (Myopes), Laser Assisted in Situ Keratomileusis (LASIK) (myopes) and LASIK (hyperopes). The preop and 3 months postop Goldmann correlated IOP (IOPg), corneal compensated IOP (IOPcc), corneal hysteresis (CH), and corneal resistance factor (CRF) were measured by the ORA. Central corneal thickness (CCT) was measured using ultrasonic pachymeter. The differences of the changes in IOPg, IOPcc, CH, CRF and CCT between the three groups were estimated. A General Linear Model was …