作者
Phillip Jonathan Buckhurst, James Stuart Wolffsohn, Sunil Shah, Shehzad Anjam Naroo, Leon Nicholas Davies, Emma Julie Berrow
发表日期
2009/7/1
期刊
British Journal of Ophthalmology
卷号
93
期号
7
页码范围
949-953
出版商
BMJ Publishing Group Ltd
简介
Background
A new commercially available optical low coherence reflectometry device (Lenstar, Haag-Streit or Allegro Biograph, Wavelight) provides high-resolution non-contact measurements of ocular biometry. The study evaluates the validity and repeatability of these measurements compared with current clinical instrumentation.
Method
Measurements were taken with the LenStar and IOLMaster on 112 patients aged 41–96 years listed for cataract surgery. A subgroup of 21 patients also had A-scan applanation ultrasonography (OcuScan) performed. Intersession repeatability of the LenStar measurements was assessed on 32 patients
Results
LenStar measurements of white-to-white were similar to the IOLMaster (average difference 0.06 (SD 0.03) D; p = 0.305); corneal curvature measurements were similar to the IOLMaster (average difference −0.04 (0.20) D; p = 0.240); anterior chamber depth measurements …
引用总数
2009201020112012201320142015201620172018201920202021202220232024211303337314131242233182022189