作者
Mehmet Koseoglu, Aysel Hur, Aysenur Atay, Serap Cuhadar
发表日期
2011/2/15
期刊
Biochemia medica
卷号
21
期号
1
页码范围
79-85
出版商
Medicinska naklada
简介
Introduction: Hemolysis is still the most common reason for rejecting samples, while reobtaining a new sample is an important problem. The aim of this study was to investigate the effects of hemolysis in different hemolysis levels for mostly used biochemical parameters to prevent unnecessary rejecti-ons.
Materials and methods: Sixteen healthy volunteers were enrolled in the study. Four hemolysis levels were constituted according to hemoglobin concentrations and they were divided into five groups: Gro-up I: 0-0.10 g/L, Group II: 0.10-0.50 g/L, Group III: 0.51-1.00 g/L, Group IV: 1.01-2.50 g/L, Group V: 2.51-4.50 g/L. Lysis was achieved by mechanical trauma.
Results: Hemolysis interference affected lactate dehydrogenase (LD) and aspartate aminotransfera-se (AST) almost at undetectable hemolysis by visual inspection (plasma hemoglobin< 0.5 g/L). Clinically meaningful variations of potassium and total bilirubin were observed in moderately hemolyzed samples (hemoglobin> 1 g/L). Alanine aminotransferase (ALT), cholesterol, gamma glutamyltransferase (GGT), and inorganic phosphate (P) concentrations were not interfered up to severely hemolyzed levels (hemoglobin: 2.5-4.5 g/L). Albumin, alkaline phosphatase (ALP), amylase, chloride, HDL-cholesterol, creatine kinase (CK), glucose, magnesium, total protein, triglycerides, un-saturated iron binding capacity (UIBC) and uric acid differences were statistically significant, but re-mained within the CLIA limits.
Conclusion: To avoid preanalytical visual inspection for hemolysis detection, improper sample rejec-tion, and/or rerun because of hemolysis, it is recommended in this study that, routine …
引用总数
20112012201320142015201620172018201920202021202220232024410121819161827353527313010
学术搜索中的文章