作者
Alberto Diaspro, Federico Federici, Mauro Robello
发表日期
2002/2/1
期刊
Applied optics
卷号
41
期号
4
页码范围
685-690
出版商
Optica Publishing Group
简介
The effects of the refractive-index mismatch in confocal laser scanning microscopy were extensively studied. The axial aberration induced in the case of fluorescent microspheres was measured. The data were used to take into account the mismatch-induced aberrations and to consider object-size influence. Then we focused on the effect of refractive-index mismatch on the effective system’s point-spread function under different mismatch conditions and on depth of focusing. We experimentally verified that the peak of the point-spread function intensity profile decreases and the point-spread function itself progressively broadens as a function of the combined effect of the refractive-index mismatch and of the penetration depth, leading to a worsening of the system’s overall performances. We also performed these same measurements by embedding subresolution beads in an oocyte’s cytoplasm, which can be …
引用总数
2003200420052006200720082009201020112012201320142015201620172018201920202021202220232024551937103124441281379136972