作者
C Bou Serhal, R Jacobs, F Gijbels, M Quirynen, D Van Steenberghe, H Bosmans, R Hermans
发表日期
2001/10
期刊
Clinical oral implants research
卷号
12
期号
5
页码范围
473-478
出版商
Munksgaard International Publishers
简介
For several radiological examinations, a clinician can select between conventional and spiral computed tomography. Using both techniques, this study aimed at evaluating the difference in absorbed doses when examining a single lateral jaw segment in a human cadaver head and Rando phantom. The present study involved the placement of thermoluminescent dosimeter (TLD) chips (GR‐200) in the thyroid gland, and bilaterally, in the parotid and submandibular glands and the lenses of the eyes in both a human cadaver and a Rando phantom at corresponding locations. Consecutive conventional spiral tomographic examinations were carried out in both the left upper and lower premolar area, using a Cranex TOME® multifunctional unit. Each examination consisted of 4 slices with a 2 mm slice thickness and exposure parameters of 57 kV, 56 seconds and 1.6–2.0 mA. Regarding spiral computed tomography (CT …
引用总数