Celia Zhou, Channa Yahathugoda, Lalindi De Silva, Upeksha Rathnapala, Grant Owen, Mirani Weerasooriya, Ramakrishna U Rao, Gary J Weil, Philip J Budge
发表日期
2019/10/4
期刊
PLoS neglected tropical diseases
卷号
13
期号
10
页码范围
e0007762
出版商
Public Library of Science
简介
Background
The Global Programme to Eliminate Lymphatic Filariasis (LF) emphasizes hygiene, exercise, and other measures to reduce morbidity and disability related to LF. We recently reported that a portable, three-dimensional, infrared imaging system (3DIS) provides accurate limb volume measurements in patients with filarial lymphedema. To assess the practical utility of repeated 3DIS measurements for longitudinal lymphedema management, we examined intraday and day-to-day leg volume changes in adults with filarial lymphedema in southern Sri Lanka.
Methodology and principal findings
We assessed 41 participants with lower extremity lymphedema (stages 1–6) in their homes in the mornings (6:00–9:00 AM) and afternoons (2:00–6:00 PM) of three days within one calendar week. Two examiners performed replicate 3DIS volume measurements at each visit. Median coefficient of variation among replicate volume measurements was 1.7% (IQR 1.1% - 2.3%) for left legs and 2.2% (IQR 1.6% - 2.8%) for right legs. Median intraday volume increase was 3.0%. Range among daily volume measurements tended to be lower for afternoon measurements (median 2.25%, IQR 1.4%– 5.4%) than for morning measurements (median 3.0%, IQR 1.4% - 8.4%).
Conclusions and significance
Limb volume measurements by 3DIS are accurate and reproducible, and this technique is feasible for use in patients’ homes. We have developed practical suggestions for optimal outcomes with 3DIS. Duplicate measurements should be performed and repeat assessments should be done at approximately the same time of day to minimize bias. Duplicate …
引用总数
201920202021202220232024111241
学术搜索中的文章