作者
Tatsuya J Arai, Joris Nofiele, Ananth J Madhuranthakam, Qing Yuan, Ivan Pedrosa, Rajiv Chopra, Amit Sawant
发表日期
2016/6
期刊
Medical physics
卷号
43
期号
6Part1
页码范围
2807-2820
出版商
American Association of Physicists in Medicine
简介
Purpose
Sparse‐sampling and reconstruction techniques represent an attractive strategy to achieve faster image acquisition speeds, while maintaining adequate spatial resolution and signal‐to‐noise ratio in rapid magnetic resonance imaging (MRI). The authors investigate the use of one such sequence, broad‐use linear acquisition speed‐up technique (kt BLAST) in monitoring tumor motion for thoracic and abdominal radiotherapy and examine the potential trade‐off between increased sparsification (to increase imaging speed) and the potential loss of “true” information due to greater reliance on a priori information.
Methods
Lung tumor motion trajectories in the superior–inferior direction, previously recorded from ten lung cancer patients, were replayed using a motion phantom module driven by an MRI‐compatible motion platform. Eppendorf test tubes filled with water which serve as fiducial markers were …