作者
Michael Jerosch‐Herold, Cory Swingen, Ravi Teja Seethamraju
发表日期
2002/5
期刊
Medical physics
卷号
29
期号
5
页码范围
886-897
出版商
American Association of Physicists in Medicine
简介
Magnetic resonance (MR) imaging during the first pass of an injected contrast agent has been used to assess myocardial perfusion, but the quantification of blood flow has been generally judged as too complex for its clinical application. This study demonstrates the feasibility of applying model‐independent deconvolution to the measured tissue residue curves to quantify myocardial perfusion. Model‐independent approaches only require minimal user interaction or expertise in modeling. Monte Carlo simulations were performed with contrast‐to‐noise ratios typical of MR myocardial perfusion studies to determine the accuracy of the resulting blood flow estimates. With a B‐spline representation of the tissue impulse response and Tikhonov regularization, the bias of blood flow estimates obtained by model‐independent deconvolution was less than 1% in all cases for peak contrast to noise ratios in the range from 15:1 …
引用总数