作者
Zoltan Sallo, Peter Perge, Bernadett Balogi, Gabor Orban, Katalin Piros, Szilvia Herczeg, Klaudia Vivien Nagy, Istvan Osztheimer, Pál Ábrahám, Bela Merkely, Laszlo Geller, Nandor Szegedi
发表日期
2022/7/7
期刊
Frontiers in Cardiovascular Medicine
卷号
9
页码范围
935705
出版商
Frontiers Media SA
简介
Introduction
High-power short-duration (HPSD) radiofrequency ablation has been proposed to produce rapid and effective lesions for pulmonary vein isolation (PVI). We aimed to evaluate the procedural characteristics and the first-pass isolation (FPI) rate of HPSD and very high-power short-duration (vHPSD) ablation compared to the low-power long-duration (LPLD) ablation technique.
Methods
One hundred fifty-six patients with atrial fibrillation (AF) were enrolled and assigned to LPLD, HPSD, or vHPSD PVI. The energy setting was 30, 50, and 90 W in the LPLD, HPSD, and vHPSD groups, respectively. In the vHPSD group, 90 W/4 s energy delivery was used in the QMODE+ setting. In the other groups, ablation index-guided applications were delivered with 30 W (LPLD) or 50 W (HPSD).
Results
Bilateral PVI was achieved in all cases. Compared to the LPLD group, the HPSD and vHPSD groups had shorter procedure time [85 (75–101) min, 79 (65–91) min, and 70 (53–83) min], left atrial dwelling time [61 (55–70) min, 53 (41–56) min, and 45 (34–52) min], total RF time [1,567 (1,366–1,761) s, 1,398 (1,021–1,711) s, and 336 (247–386) s], but higher bilateral FPI rate (57, 78, and 80%) (all p-values < 0.01). The use of HPSD (OR = 2.72, 95% CI 1.15–6.44, p = 0.023) and vHPSD (OR = 2.90, 95% CI 1.24–6.44, p = 0.014) ablation techniques were associated with a higher probability of bilateral FPI. The 9-month AF-recurrence rate was lower in case of HPSD and vHPSD compared to LPLD ablation (10, 8, and 36%, p = 0.0001). Moreover, the presence of FPI was associated with a lower AF-recurrence rate at 9-month (OR = 0.09, 95% CI 0.04–0.24 …
引用总数