简介
Objective
To compare visual, refractive, and patient-reported outcomes of patients implanted with one of three types of extended depth of focus (EDOF) intraocular lenses.
Setting
Asian Eye Institute, Philippines.
Design
Retrospective cohort study.
Methods
Subjects implanted with Symfony (Johnson and Johnson, USA), IC-8 (AcuFocus, USA), and WIOL (Medicem, Czech Republic) EDOF intraocular lenses were recruited. Spherical equivalent, uncorrected and corrected visual acuity, defocus curve, and modulation-transfer function, Strehl ratio, and visual Strehl optical transfer function values, photic phenomena, and questionnaire answers were measured and assessed.
Results
A total of 32 eyes with the Symfony lens, 30 with the IC-8 lens, and 32 with the WIOL lens were included in the study. Mean postoperative spherical equivalent was −0.24 D for the Symfony, −0.17 D for the IC-8, and 0.27 D for WIOL. There were no …