作者
Qian Wang, Zhen Liu, Sibylle I Ziegler, Kuangyu Shi
发表日期
2015/6/18
期刊
Physics in Medicine & Biology
卷号
60
期号
13
页码范围
5261
出版商
IOP Publishing
简介
Position-sensitive positron cameras using silicon pixel detectors have been applied for some preclinical and intraoperative clinical applications. However, the spatial resolution of a positron camera is limited by positron multiple scattering in the detector. An incident positron may fire a number of successive pixels on the imaging plane. It is still impossible to capture the primary fired pixel along a particle trajectory by hardware or to perceive the pixel firing sequence by direct observation. Here, we propose a novel data-driven method to improve the spatial resolution by classifying the primary pixels within the detector using support vector machine. A classification model is constructed by learning the features of positron trajectories based on Monte-Carlo simulations using Geant4. Topological and energy features of pixels fired by 18 F positrons were considered for the training and classification. After applying the …
引用总数
2016201720182019202020212022202320241112422