作者
Sina Pourang, Debnath Maji, Ujjal DS Sekhon, Anirban Sen Gupta, Michael A Suster, Pedram Mohseni
发表日期
2020/10/25
研讨会论文
2020 IEEE SENSORS
页码范围
1-4
出版商
IEEE
简介
This paper reports on the use of a microfluidic dielectric sensor, termed ClotChip, for monitoring the fibrin polymerization effects on whole blood coagulation. The ClotChip sensor readout captures the temporal variation in the normalized real part of the blood dielectric permittivity at 1MHz as it undergoes coagulation. A new readout parameter, S max , is identified in the ClotChip readout curve, which represents the maximum permittivity slope after a permittivity peak is reached. Experiments based on ex vivo treatment of whole blood samples with H-Gly-Pro-Arg-Pro-OH (GPRP) - a fibrin-polymerization inhibitor - demonstrate that S max decreases with increasing GPRP concentrations corresponding to higher levels of fibrin inhibition. Furthermore, S max is shown to exhibit a strong positive correlation (Pearson's r = 0.883, p <; 0.001, n = 12) with the α-angle readout parameter of rotational thromboelastometry - a …