Telmo Pereira, João Maldonado, Rita Coutinho, Eva Cardoso, Mafalda Laranjeiro, Ivo Andrade, Jorge Conde
发表日期
2014/10/1
期刊
Blood pressure monitoring
卷号
19
期号
5
页码范围
280-287
出版商
LWW
简介
Methods and results
This was a cross-sectional study that included 15 patients (seven women), mean age 62.07±10.59 years, referenced for cardiac catheterization. Pressure curves were obtained simultaneously in the ascending aorta (invasively) and in the right common carotid artery (using the Complior Analyse). Mean central arterial pressures, augmentation indexes, and wave morphology obtained using both methods were compared. A good concordance between methods was obtained for all the parameters measured, with intraclass correlation coefficients above 0.9. Bland–Altman analysis also indicated a good accuracy profile of the Complior device, with small mean differences observed for all parameters and most values confined within 2 SD of the mean difference. This was further confirmed by the strong Pearson correlation coefficients, with r 2 coefficients above 0.92 for all the variables studied. The …