作者
Xinhui Duan, Zhoubo Li, Lifeng Yu, Shuai Leng, Ahmed F Halaweish, Joel G Fletcher, Cynthia H McCollough
发表日期
2015/12
期刊
American journal of roentgenology
卷号
205
期号
6
页码范围
1203-1207
出版商
American Roentgen Ray Society
简介
OBJECTIVE. The purpose of this phantom study was to determine the utility of a third-generation dual-source CT scanner with increased dual-energy spectral separation in differentiating urinary stone composition.
MATERIALS AND METHODS. Eighty-seven urinary stones from humans were scanned in 35-, 40-, 45-, and 50-cm wide anthropomorphic phantoms with a third-generation dual-source scanner (system A) with a high-energy beam of 150 kV plus 0.6-mm tin filtration (Sn). The low-energy data were acquired at 70, 80, 90, and 100 kV. A second-generation dual-source scanner (system B) was used to acquire data at 140 kV plus 0.4-mm Sn for the high-energy and 80 or 100 kV for the low-energy images. Volume CT dose index was matched for a given phantom size. CT number ratios were calculated and used to differentiate uric acid from non–uric acid stones and oxalate from apatite stones in an ROC …