作者
Jia Wang, Xinhui Duan, Jodie A Christner, Shuai Leng, Lifeng Yu, Cynthia H McCollough
发表日期
2011/11
期刊
Medical physics
卷号
38
期号
11
页码范围
6084-6092
出版商
American Association of Physicists in Medicine
简介
Purpose:
The purpose of this work was to evaluate dose performance and image quality in thoracic CT using three techniques to reduce dose to the breast: bismuth shielding, organ‐based tube current modulation (TCM) and global tube current reduction.
Methods:
Semi‐anthropomorphic thorax phantoms of four different sizes (15, 30, 35, and 40 cm lateral width) were used for dose measurement and image quality assessment. Four scans were performed on each phantom using 100 or 120 kV with a clinical CT scanner: (1) reference scan; (2) scan with bismuth breast shield of an appropriate thickness; (3) scan with organ‐based TCM; and (4) scan with a global reduction in tube current chosen to match the dose reduction from bismuth shielding. Dose to the breast was measured with an ion chamber on the surface of the phantom. Image quality was evaluated by measuring the mean and standard deviation of CT …
引用总数