作者
Z Yu, S Leng, SM Jorgensen, Z Li, R Gutjahr, B Chen, X Duan, AF Halaweish, L Yu, EL Ritman, CH McCollough
发表日期
2015
期刊
Proceedings of SPIE--the International Society for Optical Engineering
卷号
9412
出版商
NIH Public Access
简介
X-ray computed tomography (CT) with energy-discriminating capabilities presents exciting opportunities for increased dose efficiency and improved material decomposition analyses. However, due to constraints imposed by the inability of photon-counting detectors (PCD) to respond accurately at high photon flux, to date there has been no clinical application of PCD-CT. Recently, our lab installed a research prototype system consisting of two x-ray sources and two corresponding detectors, one using an energy-integrating detector (EID) and the other using a PCD. In this work, we report the first third-party evaluation of this prototype CT system using both phantoms and a cadaver head. The phantom studies demonstrated several promising characteristics of the PCD sub-system, including improved longitudinal spatial resolution and reduced beam hardening artifacts, relative to the EID sub-system. More importantly …
引用总数
20152016201720182019202020212022202320243111210784341
学术搜索中的文章
Z Yu, S Leng, SM Jorgensen, Z Li, R Gutjahr, B Chen… - Proceedings of SPIE--the International Society for …, 2015