作者
Jia Wang, Xinhui Duan, Jodie A Christner, Shuai Leng, Lifeng Yu, Cynthia H McCollough
发表日期
2012/11
期刊
Medical physics
卷号
39
期号
11
页码范围
6764-6771
出版商
American Association of Physicists in Medicine
简介
Purpose
For the purpose of size‐specific dose estimation, information regarding patient attenuation is required. The purpose of this work is to describe a method for measuring patient attenuation and expressing the results in terms of a water cylinder, with cross sectional areaAw, which would absorb the same average dose as the irradiated patient. The ability to calculate Aw directly from the CT image was validated with Monte Carlo simulations and an analytical model.
Methods
A series of virtual cylinders were created with diameters ranging from 10 to 40 cm and lengths of 40 cm. The cylinders were given an atomic number equal to that of water; the density of the cylinders was varied from 0.26 to 1.2 g/cm3. The average dose to the cylinders from an axial scan at the longitudinal center position was calculated using Monte Carlo simulation and an analytical model. The relationship between phantom cross sectional …