作者
B Li, HC Lee, X Duan, C Shen, L Zhou, X Jia, M Yang
发表日期
2017/8/9
期刊
Physics in Medicine & Biology
卷号
62
期号
17
页码范围
7056
出版商
IOP Publishing
简介
The dual-energy CT-based (DECT) approach holds promise in reducing the overall uncertainty in proton stopping-power-ratio (SPR) estimation as compared to the conventional stoichiometric calibration approach. The objective of this study was to analyze the factors contributing to uncertainty in SPR estimation using the DECT-based approach and to derive a comprehensive estimate of the range uncertainty associated with SPR estimation in treatment planning. Two state-of-the-art DECT-based methods were selected and implemented on a Siemens SOMATOM Force DECT scanner. The uncertainties were first divided into five independent categories. The uncertainty associated with each category was estimated for lung, soft and bone tissues separately. A single composite uncertainty estimate was eventually determined for three tumor sites (lung, prostate and head-and-neck) by weighting the relative proportion …
引用总数
20172018201920202021202220232024113381219129
学术搜索中的文章