作者
Jia Wang, Jodie A Christner, Xinhui Duan, Shuai Leng, Lifeng Yu, Cynthia H McCollough
发表日期
2012/11
期刊
Medical physics
卷号
39
期号
11
页码范围
6772-6778
出版商
American Association of Physicists in Medicine
简介
Purpose
To estimate attenuation using cross sectional CT images and scanned projection radiograph (SPR) images in a series of thorax and abdomen phantoms.
Methods
Attenuation was quantified in terms of a water cylinder with cross sectional area ofAw from both the CT and SPR images of abdomen and thorax phantoms, where Aw is the area of a water cylinder that would absorb the same dose as the specified phantom. SPR and axial CT images were acquired using a dual‐source CT scanner operated at 120 kV in single‐source mode. To use the SPR image for estimating Aw, the pixel values of a SPR image were calibrated to physical water attenuation using a series of water phantoms. Aw and the corresponding diameter Dw were calculated using the derived attenuation‐based methods (from either CT or SPR image). Aw was also calculated using only geometrical dimensions of the phantoms (anterior …