作者
Megan C Jacobsen, Erik NK Cressman, Eric P Tamm, Dodge L Baluya, Xinhui Duan, Dianna D Cody, Dawid Schellingerhout, Rick R Layman
发表日期
2019/8
期刊
Radiology
卷号
292
期号
2
页码范围
414-419
出版商
Radiological Society of North America
简介
Background
Assessments of the quantitative limitations among the six commercially available dual-energy (DE) CT acquisition schemes used by major CT manufacturers could aid researchers looking to use iodine quantification as an imaging biomarker.
Purpose
To determine the limits of detection and quantification of DE CT in phantoms by comparing rapid peak kilovoltage switching, dual-source, split-filter, and dual-layer detector systems in six different scanners.
Materials and Methods
Seven 50-mL iohexol solutions were used, with concentrations of 0.03–2.0 mg iodine per milliliter. The solutions and water sample were scanned five times each in two phantoms (small, 20-cm diameter; large, 30 × 40-cm diameter) with six DE CT systems and a total of 10 peak kilovoltage settings or combinations. Iodine maps were created, and the mean iodine signal in each sample was recorded. The limit of blank (LOB) was …